Vicky TV-10Plus Series Digital Vickers Hardness Tester

Model: Vicky TV-10Plus
Code#: 821-240Plus
Warranty: 12 month
Delivery Time: 15-30days
Standard: JIS B -7735, ASTM E-384 and ISO 6507-2
Certificate: CE

Description

Characteristics

◆ Vickers hardness tester with colour touch screen;

◆ 8 inches colour touch screen, high resolution, easy to operate;

◆ Strong and stable body;

◆ LED illumination, light, adjustable, long life and no heat;

◆ Data storage for review and printing;

◆ Rotatable turret with five working positions;

◆ 250 lines encode with more precise measurement;

◆ Conversion between different scales;

◆ Can be developed to semi-automatically system.

Specification

Model Vicky TV-10Plus Vicky TV-30Plus Vicky TV-50Plus
Code# 821-240Plus 821-260Plus 821-280Plus
Test   Load Kgf 0.2 / 0.3 / 0.5 / 1   / 2 / 3 / 5 / 10 0.3 / 0.5 / 1 / 3 /   5 / 10 / 20 / 30 1 / 2 / 3 / 5 / 10   / 20 / 30 / 50
Nf 1.961 / 2.942 /   4.903 / 9.807 / 19.614 / 29.42 / 49.035 / 98.07 9.807 / 19.614 /   29.421 / 49.035 / 98.07 / 196.14 / 294.21 / 490.35 9.807 / 19.614 /   29.421 / 49.035 / 98.07 / 196.14 / 294.21 / 490.35
Loading Mechanism Automatic loading   and releasing method 50~100 µm/s
Dwell Time 5 ~ 99 s
Turret Mechanism Automatic
Diamond Indenter VICKERS INDENTER   (FOR HV)
Objective Normal 10X , 40X 10 X , 20X 10X
Option 40x, 20x, 5X
Measuring   Microscope Eyepiece Electronic Type 10X
Max. Measuring   length 900 µm  ( In case of 100x )    230 µm    ( In case of 400x )
Mix. Measuring unit 0.1µm (0.01µm for   CTS type)
Max.   Specimen Height 200mm
Max. Specimen Depth 160mm
Measuring Hardness   Scale HV, HK
LCD-Touch Operation   Panel Load start / set up   zero / dwell time / light adjustment / transformation scale turret location /   data processing ( browse / delete / print )System settings( test conditions /   language/Communications/factorysettings )
Numerical Display D1 value / D2 value   / HV-HK Instructions / hardness value /conversion value / load value / dwell   time / test times / turret location
Print Output Measured data and   statistic data ( Max. Value / Min. value / Mean value / Range / Standard   deviation )
Light Source LED Illuminator
Accuracy Conform to : JIS B   -7735, ASTM E-384 and ISO 6507-2
Photographic Device Capable to mount   photographic device ( Option)
Dimensions W 226 × D 560 × H   660 mm
Weight 72KG
Power Supply Single phase AC 220   V 50/60 Hz

 